DentalCare Logo

The Concepts of Four-handed Dentistry Including Ergonomic Instrument Transfer and Exercises for Stress Reduction

Course Number: 643

The Single-handed Transfer

The single handed instrument transfer with a right handed operator is illustrated in the following procedural outline.

The Single-handed Transfer - Figure 7
  • Assemble instruments in sequence of use and place the instrument tray as close to the patient as possible. The tray may be positioned in a vertical or horizontal position.

The Single-handed Transfer - Figure 8
  • If a tray or cassette is not used as in some dental specialties, prepare the materials in a logical sequence of use.

The Single-handed Transfer - Figure 9
  • Place auxiliary equipment such as the anesthetic syringe or dental dam materials on the mobile cabinet farthest from the patient.

  • At the beginning of the procedure, pick up the mirror and explorer from the tray with the

    mirror in your right hand and the explorer in your left for a right-handed operator.

The Single-handed Transfer - Figure 10
  • Simultaneously pass the mirror with the right hand and the explorer with the left hand.

The Single-handed Transfer - Figure 11
  • Pick up the instrument to be transferred in the left hand and position it between the first finger and thumb at the non-working third of the instrument.

The Single-handed Transfer - Figure 12
  • Rest the instrument on the middle finger, making certain that the working end is positioned for the correct arch and position it within 10-12 inches from the operator’s hand in readiness for a transfer when needed.

The Single-handed Transfer - Figure 13
  • The operator signals for an exchange by moving the instrument being used from the tooth and bringing it outside the mouth. When possible, a finger rest should be maintained.

The Single-handed Transfer - Figure 14
  • The assistant grasps and tucks the used instrument toward the wrist with the pick-up portion of the hand.1

The Single-handed Transfer - Figure 15
  • The new instrument is delivered into the operator’s hand with the delivery portion of the hand and the operator returns to the mouth with the new instrument.

The Single-handed Transfer - Figure 1
  • With the thumb, the assistant rolls the instrument from the palm up to the ring finger until it is above the first knuckle. Take care to avoid puncturing the gloves.

The Single-handed Transfer - Figure 2
  • Fold the index and middle fingers under the handle and return the instrument to the holding position.

The Single-handed Transfer - Figure 3
  • If the instrument is not to be used again, it can be returned to the proper position on the tray.

  • When the A/W syringe and the oral evacuator are used, the assistant places the A/W syringe in the right hand to free the other for the instrument transfer.

The Single-handed Transfer - Figure 4

Bulky instruments can be transferred in the same single-handed exchange as described above. Many dental units, such as the split unit, prohibit effective handpiece transfer since the handpiece is outside the 21-inch radius of the assistant’s hand. This unit placement leads to decreased effectiveness in terms of time and motion. When a transthorax or rear unit is used and handpieces are within the transfer zone, the assistant should then utilize the following procedure:2

  • The handpiece is made parallel with the instrument to be exchanged.

The Single-handed Transfer - Figure 6
  • The return of the handpiece in the pick-up portion of the assistant’s hand is done in the same manner as any other instrument even though it is bulkier. For this reason, the two fingers used in the pick-up method provide greater stability. This same exchange can be used when switching from one handpiece to another but may need to be done in an opposite motion to avoid hose tangling.

The Single-handed Transfer - Figure 5
  • The handpiece is then returned to the dental unit.